Installation
The TDC sensor can be operated
in either a lab or process environment. In the lab
the sensor is simply immersed into a sample contained
in a sampling tube, beaker, or jar. Process installation
involves insertion into a tank or process line. Considerations
include achieving good thermal coupling, appropriate
flow for sample refresh, and ease of access for installation
and maintenance. In some cases it is helpful to provide
a sampling port for periodic off line checks. Our
sensor is quite flexible with regard to the installation
possibilities. Several examples are illustrated below.
We will also be glad to work with you regarding your
particular needs, simply contact sales@tdcollaborative.
Bypass Piping
Installation
Orientation of the sensor is not critical. We provide
the adaptor [shown grey] to mate the sensor NPT thread
to virtually any piping of one inch or greater size.
Line pressures up to 7 bar [100psi] can be easily accommodated
with no measurable pressure effect on the sensor.
Tank Installation
A tank installation from the top, with frequent cleanup
and extraction will benefit from use of a protective
cage on the sensor [see
products/accessories]. The cage has an extension
shaft for handling and mounting. A typical mount of
this sort might be:
Side Mount Tank Installation
Alternatively, the sensor might be mounted permanently
through a side wall flange fitting with an appropriate
adaptor.
|
|